Tawfik Ahmed, Moustafa Physics Department - Faculty of Science - Mansoura University, Mansoura, Egypt , Ahmed Baker, Naer Physics Department - Faculty of Science - Mansoura University, Mansoura, Egypt , Mohammed El-Shahat, Khaled Clinical Oncology Department - Faculty of Medicine - Al Azhar University, Mansoura, Egypt , Farouk Akl, Mohamed Clinical Oncology and Nuclear Medicine Department - Faculty of Medicine - Mansoura University, Mansoura, Egypt , Shawky Shawata, Ahmed Physics Department - Faculty of Science - Mansoura University, Mansoura, Egypt
Abstract :
Background: Quality assurance for intensity modulated radiation therapy (IMRT)
depends on the type of dosimetry system and its evaluation procedure. We can check
both the intensity and distribution of each field as the required pretreatment verification
with dosimetry systems.
Method: Treatment verification for different plans (IMRT and RapidArc) applied
on localized prostate cancer patients was done with electronic portal imager device
(EPID), Delta4). The EPID used was Varian aS1000 mounted on Varian (TrueBeam)
Linac with gamma criteria set to ΔD=3% and Δd=3mm. RapidArc plans were designed
by arcs (179.0o CCW to 181.0o and 181.0o CW to 179.0o) under the same gamma
criteria of (ΔD=3%, DTA=3mm and Δ-index ≤1), while the threshold dose was 20%.
Results: Evaluation analysis is passed for IMRT prostate plans with the area
gamma <1.0 which equaled 99.1% (99.1% of the pixels had gamma<1) within a
tolerance of 95.0%, area gamma >0.8 (was equal to 2.1%) / area gamma >1.2 (was
equal to 0.3%) and the average dose difference was 0.42CU. Delta4 dosimetry system
was assessed with RapidArc plan; the agreement between the measured and planned
doses was ±1% and gamma analysis resulted in 100% data points with the same
agreement conditions.
Conclusion: Portal dosimetry provided a good verification of the treatment unit
ability to deliver doses according to plan. For an IMRT field comprised of several
subfields, it could give rise to much more errors. RapidArc plans were verified using
Delta4 system, which generated excellent dosimetry results. Periodic calibration was
recommended for Delta4 dosimetry system; radiation damage affected sensitivity by
>1% every 1kGy.
Keywords :
Radiotherapy , Intensity-modulated , RapidArc , Radiometry , EPID , Delta4