Title of article :
Readout temperature effect on the response of the genipin gel dosimeter
Author/Authors :
Bahrami, F Department of Medical Radiation Engineering - Science and Research Branch - Islamic Azad University, Tehran, Iran , Abtahi, S.M.M Physics Department - Imam Khomeini International University, Qazvin, Iran , Sardari, D Department of Medical Radiation Engineering - Science and Research Branch - Islamic Azad University, Tehran, Iran , Bakhshandeh, M Department of Radiology Technology - Allied Medical Faculty - Shahid Beheshti University of Medical Sciences, Tehran, Iran
Abstract :
Background: The present study aimed to investigate the dependence of the response of the genipin gel dosimeter on readout temperatures. Materials and Methods: To this aim, after the gel preparation, they were exposed by a Varian Medical Linac accelerator with a dose range of 0-10 Gy. Then the
irradiated samples were readout by a spectrophotometer device at 5°C, 10°C,
15°C and 20 °C temperatures. Results: Results revealed that the response of
the genipin gel dosimeter is extremely temperature-dependent. For the
mentioned readout temperatures, the maximum dose sensitivity was related
to 10°C and its value was 35×10-3cm-1Gy-1. Investigation of the AUS (area
under spectra) depicted that the maximum dose sensitivity was also related
to 10°C. Furthermore, investigation of the dose resolution of this gel
dosimeter showed that the minimum detectable dose was related to 10°C,
and its amount was 0.24 Gy. Conclusion: The results confirmed a noticeable
change in the response of the genipin gel dosimeter due to variations of the
readout temperature. It can be concluded that the response sensitivity of this radiochromic gel dosimeter increased by readout temperature decreasing.
Keywords :
Genipin gel dosimeter , dose sensitivity , readout temperature , dose resolution
Journal title :
International Journal of Radiation Research