Title of article :
Polycyclic Aromatic Hydrocarbons from Bottled Water: A New Automated Solid Phase Extraction Method and Simulation of Cancer Risk
Author/Authors :
Mahdavi Doost, Fatemeh Department of Environmental Health Engineering - Health Faculty - Qazvin University of Medical Sciences, Qazvin, Iran , Emamjome, Mohammad Mehdi Department of Environmental Health Engineering - Health Faculty - Qazvin University of Medical Sciences, Qazvin, Iran , Ali Jamali, Hamze Department of Environmental Health Engineering - Health Faculty - Qazvin University of Medical Sciences, Qazvin, Iran , Karyab, Hamid Research Center for Social Determinants of Health - Qazvin University of Medical Sciences, Qazvin, Iran
Abstract :
Considering the adverse effects of polycyclic aromatic hydrocarbons (PAHs) on human health and the
complexity of their detection in water resources, this study was developed to assess the performance of a
new automated device for the identification of PAHs in water based on the solid-phase extraction and to
simulate the cancer risk of 16 priority PAHs by Monte Carlo technique. All operational modes of extraction
were automatically performed using the proprietary software program. Three spiked aliquots of PAHs
including 100, 500, and 1000 ng/L were used to evaluate the performance of the automated-solid phase
extraction (SPE) apparatus. The time of extraction in the automatic-SPE apparatus was 50 ± 4 minutes for
simultaneous extraction of 4 water samples, which was four times faster than that of manual-SPE apparatus.
The mean recoveries of PAHs were 89.22 ± 4.94, 91.70 ± 4.45, and 94.61 ± 6.28% in spiked samples, with
a mean of 91.84 ± 5.22%. Except for naphthalene, all obtained recoveries were in an acceptable range
(85-115%). The results of Monte Carlo simulation showed that the cancer risk attributable to eight detected
PAHs including naphthalene, fluorene, phenanthrene (Phe), benzoanthracene, chrysene, benzo(k)
fluoranthene, indeno(cd)pyrene, and dibenzo(a,h)anthracene ranged from 0.05E-6 to 0.11E-6, with the
cancer risk of 0.012E-6 having the highest probability (P = 0.82). Additionally, this simulation showed
that 99% of the probability density of cancer risk was located within the range lower than 0.05E-6. Our
results showed that the novel automated-SPE apparatus could be utilized for the extraction of PAHs from
water resources with a good recovery (85%-115%), high operational speed, and potential for separation of solvents from the air in the drying step. Therefore, this device can be used for the extraction of trace PAHs and similar organic compounds from water resources.
Keywords :
Drinking water , Polycyclic aromatic hydrocarbons , Automated solid-phase extraction , Risk , Monte Carlo method
Journal title :
Avicenna Journal of Environmental Health Engineering