Title of article
Design of a two-dimensional pseudo coincidence Compton suppressor system for neutron activation analysis
Author/Authors
Biganeh, Ali Physics & Accelerators Research School - Nuclear Science and Technology Research Institute, Tehran, Iran , Shakeri Jooybari, Banin Physics & Accelerators Research School - Nuclear Science and Technology Research Institute, Tehran, Iran
Pages
5
From page
17
To page
21
Abstract
Compton scattering events are the main source of error on the peak counting during the
Neutron Activation Analysis (NAA). The Compton suppressor system in instrumental
NAA reduces the detection limit of the technique and leads to a data with a higher degree
of precision. In this paper, a two-dimensional coincidence Compton suppressor system is
presented for the NAA technique. The system is established based on a CAEN digitizer
which directly records the pre-amplifier output signals of the two HPGe detectors. The
recorded events in the list mode file are analyzed offline by a Matlab code and the
correlated photopeak events are realized. The performance of the system for Compton
suppression is tested by measuring the gamma lines of Ba-133 and Cs-137 standard
sources. The results show that the presented technique provides the peak to Compton
ratio up to 104 and can be an alternative for conventional Compton suppressor systems.
Keywords
Compton suppressor , Neutron Activation Analysis , List mode , Digitizer
Journal title
Radiation Physics and Engineering
Serial Year
2022
Record number
2732662
Link To Document