• Title of article

    CTV to PTV Margins Based On CBCT Method for Prostate Cancer of Patients Treated With VMAT Technique

  • Author/Authors

    Erraoudi ، Morad Laboratory of high energy physics Modelisation Simulation - Faculty of Sciences - University Mohammed V , Bouzekraoui ، Youssef Laboratory of Sciences and Health Technologies - High Institute of Health Sciences - Hassan First University of Settat , Asmi ، Hicham Laboratory of high energy physics Modelisation Simulation - Faculty of Sciences - University Mohammed V , BENTAYEB ، FARIDA Laboratory of High Energy Physics - Faculty of Science - Mohammed V Agdal University

  • From page
    266
  • To page
    272
  • Abstract
    Introduction: Patient repositioning in treatment radiotherapy is one of the main factors that increase error of target irradiation. However additional margin is necessary to consider the uncertainties created along and around X, Y and Z-axis.Material and Methods: Set-up and random errors were calculated in translational and rotational axis for a sample of 20 prostatic patients; using daily IGRT-CBCT method. The aim of this study was to determine the additional margin that should be added from clinical target volume (CTV) to prevent toxicity and increase the irradiation precision in radiotherapy. The van Hark formula (PTV margin =2.5Σ +0.7σ) was used for all patients to perform PTV margin for prostatic localization.Results: The research performed for a sample of 20 consecutive patients. With respect to systematic error along the lateral axis, longitudinal and anterior-posterior was 2.32, 2.42 and 3.54 respectively. The Random error was 1.82, 2.19 and 1.76° along lateral axis, longitudinal and anterior-posterior respectively. The rotational systematic error was 1.49, 2.04 and 2.14° around lateral, longitudinal and anterior-posterior axis respectively. The Random error was 1.78, 1.75 and 1.63° around lateral, longitudinal and anterior-posterior axis respectively. The calculated safety margin to cover clinical target volume (CTV) taking the prostate variability into account measured 7.55, 8.08 and 10.79 mm for lateral, longitudinal and anterior posterior respectively and 7mm would be enough in the posterior side. Rotational set-up errors for almost 95% of patients were between -2° and 2°.Conclusion: The calculated safety margin in all direction was smaller than 1 cm except in anterior side that was 1 cm or more.
  • Keywords
    CTV , PTV , CBCT method , VMAT , Prostate
  • Journal title
    Iranian Journal of Medical Physics (IJMP)
  • Journal title
    Iranian Journal of Medical Physics (IJMP)
  • Record number

    2754418