Title of article
Malposition of Central Venous Pressure Catheter as a Rare Cause of Tricuspid Valve Prosthesis Malfunction: A Case Report malfunction; a case report
Author/Authors
Behjati ، Mohaddeseh Rajaie Cardiovascular Medical and Research Center - Iran University of Medical Sciences , Tohidi ، Fatemeh Department of Medical Biotechnology - School of Medicine, Iran. AND Cancer Research Center, Health Research Institute - Babol University of Medical Sciences , Baharestani ، Bahador Rajaie Cardiovascular Medical and Research Center - Iran University of Medical Sciences
From page
770
To page
772
Abstract
Prosthetic valve malfunction is a hazardous clinical entity that needs early diagnosis and management. Hereby, we present a case of tricuspid valve (TV) malfunction due to interaction of central venous pressure (CVP) catheter with TV leaflets. Our case was a 38-year-old male who referred to our center with the diagnosis of TV malfunction. Confirming the diagnosis of TV malfunction, intravenous Heparin was initiated. Anticoagulation therapy was discontinued early due to the development of significant headache in the setting of subarachnoid hemorrhage. Due to contraindication of anticoagulation therapy, he underwent redo TV. On the 4th post-operative day, echocardiography examination illustrated significant increase in gradients of TV prosthesis. At this time, fluoroscopy depicted interaction of CVP catheter with TV leaflets due to its cardiac advancement resulted in fixed TV leaflets in semi-open position. By extraction of CVP catheter the TV gradients improved immediately. Malposition of CVP catheter or large CVP catheter could be considered as one of the rare causes of TV malfunction. Repositioning of CVP catheter brings dramatic improvement in this case.
Keywords
Tricuspid valve , Prosthetic malfunction , Central venous pressure catheter
Journal title
Acta Medica Iranica
Journal title
Acta Medica Iranica
Record number
2771155
Link To Document