Title of article :
DETERMINATION OF LEVONORGESTREL IN HUMAN SERUM BY LIQUID CHROMATOGRAPHIC-ELECTROSPRAY TANDEM MASS SPECTROMETRY
Author/Authors :
Li، Guang نويسنده , , Wang، Qinggang نويسنده , , Wu، Zhuping نويسنده , , Wang، Yiming نويسنده , , Luo، Guoan نويسنده , , Wu، Erruo نويسنده , , Gao، Xuefeng نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
-102
From page :
103
To page :
0
Abstract :
A rapid, sensitive, and specific high-performance liquid chromatography electrospray tandem mass spectrometric method has been developed for the determination of levonorgestrel in human serum using norethindrone as an internal standard. Levonorgestrel was detected using multiple reaction monitoring mode after extracted from human serum by an ether extraction procedure. The calibration curve was linear over the range of 5–500 ng/mL (r2 = 0.999) with the limitation of detection of 1 ng/mL. The intra-day precision and accuracy, expressed as CV and RE, ranged from 0.5% to 10.1% and -0.2% to 2.2%. The inter-day precision and accuracy ranged from 2.1% to 13.6% and -2.3% to 3.3%. The mean recovery was 89.5% for levonorgestrel, and 90.2% for the internal standard.
Journal title :
Analytical Letters
Serial Year :
2001
Journal title :
Analytical Letters
Record number :
36666
Link To Document :
بازگشت