Title of article
Very-Large-Volume Sampling of Water in Gas Chromatography Using the Through Oven Transfer Adsorption Desorption (TOTAD) Interface for Pesticide-Residue Analysis
Author/Authors
Alario، J. نويسنده , , Perez، M. نويسنده , , Vazquez، A. نويسنده , , Villén[1]، J. نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2001
Pages
-64
From page
65
To page
0
Abstract
The Through Oven Transfer Adsorption Desorption (TOTAD) interface is used to directly introduce large volumes of water (1 mL or more) into a capillary gas chromatograph. The TOTAD interface is a greatly modified programmed temperature vaporizer injector incorporating changes that affect the pneumatics, sample introduction, solvent elimination, and operation mode. The system can easily be automated. The technique is applied to the analysis of pesticide residue in standard solutions and real water samples from the Ebro River (northeastern Spain). The speed of sample introduction was 1 mL/min, and the solvent elimination was almost complete. A nitrogen phosphorous detector is used, and the relative standard deviation varied from 5.7% to 11.7% for the absolute peak areas. The sensitivity achieved by introducing 1 mL of the sample is sufficient for most pesticide-residue analyses in water. The limits of detection ranged from 0.5 to 8.1 ng/L.
Keywords
Organic compounds , Chemical synthesis , Electronic paramagnetic resonance (EPR) , Infrared spectroscopy , Fullerenes