• Title of article

    Very-Large-Volume Sampling of Water in Gas Chromatography Using the Through Oven Transfer Adsorption Desorption (TOTAD) Interface for Pesticide-Residue Analysis

  • Author/Authors

    Alario، J. نويسنده , , Perez، M. نويسنده , , Vazquez، A. نويسنده , , Villén[1]، J. نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2001
  • Pages
    -64
  • From page
    65
  • To page
    0
  • Abstract
    The Through Oven Transfer Adsorption Desorption (TOTAD) interface is used to directly introduce large volumes of water (1 mL or more) into a capillary gas chromatograph. The TOTAD interface is a greatly modified programmed temperature vaporizer injector incorporating changes that affect the pneumatics, sample introduction, solvent elimination, and operation mode. The system can easily be automated. The technique is applied to the analysis of pesticide residue in standard solutions and real water samples from the Ebro River (northeastern Spain). The speed of sample introduction was 1 mL/min, and the solvent elimination was almost complete. A nitrogen phosphorous detector is used, and the relative standard deviation varied from 5.7% to 11.7% for the absolute peak areas. The sensitivity achieved by introducing 1 mL of the sample is sufficient for most pesticide-residue analyses in water. The limits of detection ranged from 0.5 to 8.1 ng/L.
  • Keywords
    Organic compounds , Chemical synthesis , Electronic paramagnetic resonance (EPR) , Infrared spectroscopy , Fullerenes
  • Journal title
    JOURNAL OF CHROMATOGRAPHIC SCIENCE
  • Serial Year
    2001
  • Journal title
    JOURNAL OF CHROMATOGRAPHIC SCIENCE
  • Record number

    37747