• Title of article

    A differential mobility analyzer and a Faraday cup electrometer for operation at 200-930 Pa pressure

  • Author/Authors

    al.، Kwang Soo Seol et نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2000
  • Pages
    -1388
  • From page
    1389
  • To page
    0
  • Abstract
    We have developed a differential mobility analyzer (DMA) based on the DMA devised by Seto et al. (1997) and a Faraday cup electrometer for measurement of nanometer-sized particles at a few hundred Pa and examined the operating characteristics of the DMA using the tandem DMA technique. The tandem DMA calibration establishes that the DMA successfully classifies particles in the 200-930 Pa pressure range. It was also found that the transfer function of the DMA follows the triangular transfer function and the resolution of the DMA is close to that given for an ideal case. As a standard of a minimum pressure that may be probed with the present DMA system, 400 Pa is estimated when the DMA operates with a 3nlmin^-1 sheath flow and a 1nlmin^-1 aerosol flow rate.
  • Keywords
    Epstein-Barr virus , Richters transformation
  • Journal title
    Journal of Aerosol Science
  • Serial Year
    2000
  • Journal title
    Journal of Aerosol Science
  • Record number