Radiation effects on the resolution (MTF) of the scintillator coupled CMOS APS array imager for non-destructive test X-ray imaging
Author/Authors
Kwang Huyn Kim، نويسنده , , Gyuseong Cho، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
7
From page
805
To page
811
Abstract
In this paper, the effects of the transmitted X-ray through the scintillator on the resolution in the scintillator coupled CMOS APS array imager were investigated using a modulation transfer function (MTF). By measuring the system MTF of the detector at each cumulative exposure, we found out that the resolution of the detector was not permanent but changed with the amount of the cumulative exposure into the detector. These measurements were performed under the condition of non-destructive test (NDT). The detector used in this work was two-dimensional array CMOS APS of RadEye1™ coupled with Lanex™ screen. It was possible to explain the changed MTF with the results of the dark signal increase in the sensor.