Title of article :
A Contactless CCD Dilatometer for Foil Materials
Author/Authors :
H. Wang، نويسنده , , Y. Y. Zhang، نويسنده , , G. H. He، نويسنده , , B. Q. Wang and B. L. Zhou ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1999
Abstract :
A contactless, compact, low-cost dilatometer based on a laser-pulse thermalconductivity
apparatus has been developed to measure the linear thermal expansion
of foil materials. The two sample-edge images are projected onto the array
of a charge-coupled device (CCD). Changes in sample length are determined
from measurements of the corresponding displacements of the sample-edge
images focused on the CCD. The dilalometer performance was tested by comparing
results of measurements of the thermal expansion for pure copper with
published data. The linear thermal expansion of an L-16-type foil of 20-1mm
thickness, which is a candidate material for thermocontrol layers (in engineering),
was measured with the apparatus.
Keywords :
CCD , contactless , dilatometer , FOIL , Thermal expansion.
Journal title :
International Journal of Thermophysics
Journal title :