Title of article :
A Transient Hot-Wire Instrument for the Measurement of the Thermal Conductivity of Solids up to 590 K
Author/Authors :
M. J. Assael and K. Gialou ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
A novel application of the transient hot-wire technique for measurements of the
thermal conductivity of solids up to 590 K is described. The method makes use
of a soft silicone material between the hot wires and the solid of iInterest. Measurement
of the transient temperature rise of the wires in response to electrical
step heating over a period of 20 ms to 20 s allows an absolute determination of
the thermal conductivity of the solid, as well as of the silicone paste. The
method is based on a full theoretical model with equations solved by a finiteelement
method applied to the exact geometry. Two sets of thermal-conductivity
measurements up to 590 K, employing different silicone pastes and samples of
Pyroceram 9606, are reported. At the 95% confidence level, the standard deviation
of the thermal conductivity measurements is less than 1.5%.