The newly developed quasi-steady state (QSS) method to measure the
thermal conductivity combines characteristic advantages of transient and
steady-state techniques but avoids their major drawbacks. Based upon a transient
hot strip setup, the QSS technique can be realized by adding only two
temperature sensors at different radial distances from the strip. After a short
settling time, the QSS output signal which is the measure for the thermal
conductivity is constant in time as it is for steady-state instruments. Moreover,
in contrast to transient techniques, the QSS signal is not altered by
homogeneous boundary conditions. Thus, there is no need to locate a time
window as has to be done with the transient hot wire or transient hot strip
techniques. This paper describes the assessment of the QSS standard uncertainty
of thermal conductivity according to the corresponding ISO Guide. As
has already been done in previous papers on the uncertainty of the transient
hot wire and transient hot strip techniques, first, the most significant
sources of error are analyzed and numerically evaluated. Then the results are
combined to yield an estimated overall uncertainty of 3.8%. Simultaneously,
the present assessment is used as an aid in planning an experiment and in
designing a QSS sensor to achieve minimal uncertainty. Such a sensor is used
to verify the above mentioned standard uncertainty from a run on the candidate
reference material polymethyl methacrylate.