Title of article :
A New High-Temperature Oscillating Cup Viscometer
Author/Authors :
W. Hoyer and M. Kehr ، نويسنده , , W. Hoyer and I. Egry ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
9
From page :
1017
To page :
1025
Abstract :
A new oscillating cup viscometer for temperatures up to 2,300◦C has been constructed. A vacuum furnace with a graphite heater is used for heating the sample. The temperatures of the furnace and sample are measured by both a thermocouple and a pyrometer. The temperature is controlled with a stability better than 1K. The oscillation of the cup is measured with a reflected laser beam using a position sensitive detector. The measured values of angle and time are then fitted to an analytical oscillation function. From the parameters of this function, the viscosity values are calculated using the Roscoe formalism. Measurements were carried out on pure metals at temperatures up to 1,700◦C because of limitations of the thermocouple. The obtained viscosity values showed good agreement with literature data.
Keywords :
High temperature · Oscillating cup viscometer · Pure metals · Viscosity
Journal title :
International Journal of Thermophysics
Serial Year :
2007
Journal title :
International Journal of Thermophysics
Record number :
427492
Link To Document :
بازگشت