Title of article
Measurement of Aerosol Absorption Coefficient from Teflon Filters Using Integrating Plate and Integrating Sphere Techniques
Author/Authors
Campbell، نويسنده , , Dave; Copeland، نويسنده , , Scott; Cahill، نويسنده , , Thomas ، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1995
Pages
6
From page
287
To page
292
Abstract
Optical transmission of aerosol samples collected on
Teflon filters measured by the integrating plate method
(IPM) has been used as a means of determining the
aerosol absorption coefficient, bap ʹ by the Air Quality
Group at Crocker Nuclear Lab for over 9 years. The
results of an intensive quality assurance program, using
an integrating sphere photometer configured for
measurement of the transmission of filters, demon-strate that the (PM is highly precise (± 1%) and does
not sufTer from inherent overestimation of particle
absorption due to internal scattering, a problem commonly
attributed to (PM measurements of aerosols
collected on Nuclepore filters. If the appropriate correction
is applied when filter loading exceeds a monolayer
of particles, bap can be accurately determined by
the (PM for aerosol samples collected on Teflon filters.