Title of article :
Asymmetric Instrument Response Resulting from Mixing Effects in Accelerated DMA-CPC Measurements
Author/Authors :
Russell، نويسنده , , Lynn M.; Flagan، نويسنده , , Richard C.; Seinfeld، نويسنده , , John H، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Abstract :
Several method have been proposed to accelerate the
measurements made with differential mobility analyzers
(DMA), including the Scanning Electrical Mobility
Spectrometer (SEMS) and the Scanning Mobility Particle
Sizer (SMPS). Wang and FJagan (1990) developed
a data analysis procedure that accounts for the migration
of the particles through a time-varying electric
field and the delay associated with transport from the
analyzer column outlet to the detection point. Experience
using a variety of detectors and scan rates has
indicated that the instrument response depends on the
plumbing configuration if a condensation particle
counter (CPC) is used as a detector. When a sharply
peaked distribution is analyzed, the apparent breadth
of the measured distribution depends on the detector
used and on the scan rate. Size distributions measured
with an increasing voltage scan exhibit a tail toward large particle sizes, while decreasing voltage scans produce
a more pronounced tail on the small particle end
of the distribution. The reduction in peak height with
increasing scan rate can be attributed to particle retention
in the plumbing between tbe outlet of the DMA
analyzer column and the point in the CPC where the
particles are detected optically. This paper examines
the smearing of the transfer function of the SEMS as a
result of flow non-idealities in the system. A model has
been developed to predict the distortion of the transfer
function in terms of the particle residence time distribution
within the instrument. Results of this model are
compared with calibration data as a function of the
detector employed and the scan rate. We include laboratory
observations of the phenomenon and examine
the use of data inversion techniques to retrieve the true
size distribution
Journal title :
Aerosol Science and Technology
Journal title :
Aerosol Science and Technology