Intercomparison Study of the Size-Dependent Counting Efficiency of 26 Condensation Particle Counters
Author/Authors
Wiedensohlet، نويسنده , , A.; Orsini، نويسنده , , D.; Covert، نويسنده , , D. S.; Coffmann، نويسنده , , D.; Cantrell، نويسنده , , W.; Havlicek، نويسنده , , M.; Brechtel، نويسنده , , F. J.; et. Al، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1997
Pages
19
From page
224
To page
242
Abstract
Particle detection efficiency curves for 26 condensation particle counters
were determined during a calibration workshop in preparation for the Aerosol
Characterization Experiment 1 (ACEt). Three different types of commercially
available particle counters, the ultrafine condensation particle counter (TSI-302S)
and the condensation particle counters (TSI-3010 and TSI-3760 or TSI-761O) were
investigated at default temperature and flow settings as well as for other flow rates
and temperature differences between the saturator and the condenser. Furthermore,
the pulse-height-analysis ultrafine condensation particle counter and a
TSI-3010 modified to achieve a higher temperature difference were calibrated. In
this study, the large number of particle counters investigated provided the opportunity
to obtain a more statistically significant picture of the performance of different
particle counters for different operating conditions