• Title of article

    Effect of Clean Air Core Geometry on Fine Particle Contamination And Calibration of a Virtual Impactor

  • Author/Authors

    Li، نويسنده , , Shou-Nan; Lundgren، نويسنده , , Dale A، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    11
  • From page
    625
  • To page
    635
  • Abstract
    The optimum diameter and position of the clean air core in a clean air core virtual impactor were experimentally determined. It was found that the clean air core diameter should be at least twice as large as the converging nozzle diameter and that the clean air core should be positioned so that the ratio of the clean air How velocity to the aerosol How velocity ranges from 1.5 to 5.0 at the outlet of the clean air tube. Use of a laminator in a virtual impactor was found to increase the particle loss while maintaining almost the same collection efficiency curve.
  • Journal title
    Aerosol Science and Technology
  • Serial Year
    1997
  • Journal title
    Aerosol Science and Technology
  • Record number

    430435