Title of article :
Performance of Vienna Type Differential Mobility Analyzer at 1.2-20 Nanometer
Author/Authors :
Thomas Reischl، نويسنده , , G. P.; Makela، نويسنده , , J. M.; Necid، نويسنده , , J، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1997
Abstract :
The transfer functions of the so called "Vienna type" differential
mobility analyzer (DMA) were studied experimentally at the size range of 1.2-20
om following the procedure suggested by Stolzenburg (1988) Ph.D. Thesis. For the
study, two identical DMAs were used as a tandem DMA (TDMA) arrangement to
measure nanoparticles from a hot wire wax generator. The approximate analytical
formulae by Stolzenburg describing diffusion broadening of the DMA transfer
function were observed to be valid with high accuracy from 20 nm down to - 2 nm.
In the size range of 1.2 -2 nm nominal mobility equivalent diameter small deviations
of the shape of the TDMA output signal predicted by Stolzenburg were found.
Journal title :
Aerosol Science and Technology
Journal title :
Aerosol Science and Technology