• Title of article

    Performance of Vienna Type Differential Mobility Analyzer at 1.2-20 Nanometer

  • Author/Authors

    Thomas Reischl، نويسنده , , G. P.; Makela، نويسنده , , J. M.; Necid، نويسنده , , J، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1997
  • Pages
    22
  • From page
    651
  • To page
    672
  • Abstract
    The transfer functions of the so called "Vienna type" differential mobility analyzer (DMA) were studied experimentally at the size range of 1.2-20 om following the procedure suggested by Stolzenburg (1988) Ph.D. Thesis. For the study, two identical DMAs were used as a tandem DMA (TDMA) arrangement to measure nanoparticles from a hot wire wax generator. The approximate analytical formulae by Stolzenburg describing diffusion broadening of the DMA transfer function were observed to be valid with high accuracy from 20 nm down to - 2 nm. In the size range of 1.2 -2 nm nominal mobility equivalent diameter small deviations of the shape of the TDMA output signal predicted by Stolzenburg were found.
  • Journal title
    Aerosol Science and Technology
  • Serial Year
    1997
  • Journal title
    Aerosol Science and Technology
  • Record number

    430437