Title of article
Development of a hybrid battery system for an implantable biomedical device, especially a defibrillator/cardioverter (ICD)
Author/Authors
Jürgen Drews، نويسنده , , R. Wolf، نويسنده , , G. Fehrmann، نويسنده , , R. Staub، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1999
Pages
5
From page
107
To page
111
Abstract
An implantable defibrillator battery has to provide pulse power capabilities as well as high energy density. Low self-discharge rates are mandatory and a way to check the remaining available capacity is necessary. These requirements are accomplished by a system consisting of a lithium/manganese dioxide 6 V battery, plus a lithium/iodine-cell. The use of a high rate 6 V double-cell design in combination with a high energy density cell reduces the total volume required by the power source within an implantable defibrillator. The design features and performance data of the hybrid system are described.
Keywords
Applications/medical , Lithium/manganese dioxide batteries , Lithium/iodine batteries