Title of article :
The use of an implantable force transducer to measure patellar tendon forces in goats
Author/Authors :
Donna L. Korvick، نويسنده , , John F. Cummings، نويسنده , , Edward S. Grood، نويسنده , , John P. Holden، نويسنده , , Seth M. Feder، نويسنده , , David L. Butler، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
5
From page :
557
To page :
561
Abstract :
Patellar tendon (PT) force was measured during activity with an implantable force transducer (IFT) in adult goats. PT force, vertical ground reaction force (VGRF) and the animalʹs speed were recorded for standing, walking and trotting. Following data collection, animals were euthanized and the IFT calibrated in vitro. Standing PT force averaged 207 N. Maximum PT force was approximately 800 N for walking and 1000 N for trotting and occurred at mid-stance. PT force dropped from 200 N at toe-off to 0 N by mid-swing. For each activity, the PT force increased with increases in VGRF. Maximum in vivo PT stress occurred during trotting and measured 29 MPa. This study demonstrates the IFTʹs usefulness in measuring tendon force directly.
Keywords :
Tendon , Goats/animal , biomechanics , Force measurements , Patella tendon. , Transducers
Journal title :
Journal of Biomechanics
Serial Year :
1996
Journal title :
Journal of Biomechanics
Record number :
450297
Link To Document :
بازگشت