Title of article :
Instrumented implant for measuring tibiofemoral forces
Author/Authors :
Kenton R. Kaufman، نويسنده , , Nebojsa Kovacevic، نويسنده , , Steven E. Irby، نويسنده , , Clifford W. Colwell، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
5
From page :
667
To page :
671
Abstract :
Experimental measurement of loads occurring in the human knee joint will allow validation of analytical models and provide data for the design of total knee implants. A customized transducer was developed to measure the dynamic tibiofemoral force and center of pressure after total knee arthroplasty. The transducer consists of a standard tibial component to which four uniaxial load cells and an additional tibial tray have been added. The transducer was calibrated using a loading device traceable to the National Institute of Standards and Technology (NIST). The transducer was accurate to within 1% in magnitude, 0.07 mm in medial/lateral location and 0.24 m in anterior/posterior location.
Keywords :
Force transducer , Knee , Tibiofemoral force , Total joint arthroplasty