Title of article :
Evaluation of a long-range transmitter for use with a magnetic tracking device in motion analysis
Author/Authors :
Judd S. Day، نويسنده , , Genevieve A. Dumas، نويسنده , , Duncan J. Murdoch، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Abstract :
The suitability of a long-range transmitter was evaluated for use with a Polhemus Fastrak® magnetic tracking device in kinesiologic studies. Performance was judged by comparing positional and rotational accuracy to a standard transmitter. Data were obtained at distances of up to 2.7 and 5.0 m for the standard and long-range transmitters, respectively. Use of the long-range transmitter improved system performance allowing reproducible measurements at a greater distance. However, it is necessary to calibrate the system in each new test environment as there can be significant distortion of the magnetic field.
Keywords :
Magnetic tracking device , Error , Kinematic , three-dimensional , accuracy
Journal title :
Journal of Biomechanics