• Title of article

    Improved RSA accuracy with DLT and balanced calibration marker distributions with an assessment of initial-calibration

  • Author/Authors

    Anthony M. T. Choo، نويسنده , , Thomas R. Oxland، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    259
  • To page
    264
  • Abstract
    Roentgen stereophotogrammetric analysis (RSA) has been used for over 25 years for accurate micromotion measurement in a wide variety of orthopaedic applications. This study investigated two possible improvements to the method. First, direct linear transformation (DLT) was compared to the traditional RSA reconstruction algorithm. The two methods were considered with respect to standard extrapolation and interpolation calibration cages. Matlab simulations showed that reconstruction accuracy was greatly improved (>60%) by combining DLT with an even distribution of enclosing calibration markers. Second, a benchtop study using phantoms translated at 0.0254-mm intervals showed initial-calibration, followed by removal of the interpolation cage for subsequent exposures, was potentially twice as accurate as self-calibration with an extrapolation cage. These results showed optimizations for the application of RSA when unobstructed space is required.
  • Keywords
    DLT , Marker distribution , calibration , RSA , accuracy
  • Journal title
    Journal of Biomechanics
  • Serial Year
    2003
  • Journal title
    Journal of Biomechanics
  • Record number