Title of article :
A computational fluid-structure interaction analysis of a fiber-reinforced stentless aortic valve
Author/Authors :
J. de Hart، نويسنده , , F. P. T. Baaijens، نويسنده , , G. W. M. Peters، نويسنده , , P. J. G. Schreurs ، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
14
From page :
699
To page :
712
Abstract :
The importance of the aortic root compliance in the aortic valve performance has most frequently been ignored in computational valve modeling, although it has a significant contribution to the functionality of the valve. Aortic root aneurysm or (calcific) stiffening severely affects the aortic valve behavior and, consequently, the cardiovascular regulation. The compromised mechanical and hemodynamical performance of the valve are difficult to study both ‘in vivo’ and ‘in vitro’. Computational analysis of the valve enables a study on system responses that are difficult to obtain otherwise. In this paper a numerical model of a fiber-reinforced stentless aortic valve is presented. In the computational evaluation of its clinical functioning the interaction of the valve with the blood is essential. Hence, the blood–tissue interaction is incorporated in the model using a combined fictitious domain/arbitrary Lagrange–Euler formulation, which is integrated within the Galerkin finite element method. The model can serve as a diagnostic tool for clinical purposes and as a design tool for improving existing valve prostheses or developing new concepts. Structural mechanical and fluid dynamical aspects are analyzed during the systolic course of the cardiac cycle. Results show that aortic root compliance largely influences the valve opening and closing configurations. Stresses in the delicate parts of the leaflets are substantially reduced if fiber-reinforcement is applied and the aortic root is able to expand.
Keywords :
Finite element method , Fiber-reinforcement , Fluid–structure interaction , ArbitraryLagrange–Euler method , Fictitious domain method , aortic valve
Journal title :
Journal of Biomechanics
Serial Year :
2003
Journal title :
Journal of Biomechanics
Record number :
451522
Link To Document :
بازگشت