• Title of article

    Effects of helium–oxygen mixtures on endotracheal tubes: an in vitro study

  • Author/Authors

    Patrick Gerbeaux، نويسنده , , Marc Gainnier، نويسنده , , Jean-Michel Arnal، نويسنده , , Laurent Papazian، نويسنده , , Philippe Jean، نويسنده , , Jean-Marie Sainty، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    5
  • From page
    33
  • To page
    37
  • Abstract
    Question: To determine flow pattern and critical Reynolds numbers in endotracheal tubes submitted to different helium–oxygen mixtures under laboratory conditions. Materials and methods: Flow–pressure relationships were performed for seven endotracheal tubes (rectilinear position, entry length applied) with distal end open to atmosphere (predicted internal diameters: 6–9 mm). Nine helium–oxygen mixtures were tested, with FIHe varying from zero to 0.78 (increment: 10%). Nine flows were tested, with rates varying from 0.25 to 1.60 l s−1 (increment: 0.15 l s−1). Gas flow resistance was calculated, and for each endotracheal tube, a Moody diagram was realised. Flow regime and critical Reynolds numbers were then determined (fully established laminar, nonestablished laminar, smooth turbulent, or rough). Results: Even low concentration of helium in inspiratory mixture reduces endotracheal tubes resistance. Effect is maximal for high flows, small tube and high FIHe. Critical Reynolds numbers are inversely correlated to tube diameter. Answer: Under laboratory conditions, flow pattern in endotracheal tubes varies from fully established laminar to rough. Knowledge of the critical Reynolds numbers allows correct application of fluid mechanic formula when studying tube or gaseous mixture effects on respiratory mechanisms.
  • Keywords
    Helium–oxygen mixture , Heliox , Fluid-mechanics laws , Mechanical ventilation , Endotracheal tube , pressure drop
  • Journal title
    Journal of Biomechanics
  • Serial Year
    2005
  • Journal title
    Journal of Biomechanics
  • Record number

    451912