Title of article :
In vitro fatigue behaviour of a cemented acetabular reconstruction
Author/Authors :
Phillip Heaton-Adegbile، نويسنده , , Nikolaus P. Zant، نويسنده , , Jie Tong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
In this study, a hemi-pelvis of composite sawbone was implanted with a Charnley cup using a conventional bone cement and the acetabular replacement was tested under constant amplitude cyclic loads, simulating the maximum hip contact force during normal walking. The damage development in the reconstruction was detected and monitored using CT scanning at regular test intervals, verified by microscopy post testing. Three identical experimental results showed that extensive debonding at the bone–cement interface occurred around the dome region after 20 million cycles.
Keywords :
Cemented fixation , CT scans , Implant loosening , Fatigue damage
Journal title :
Journal of Biomechanics
Journal title :
Journal of Biomechanics