Title of article
Total hip arthroplasty wear simulation using the boundary element method
Author/Authors
G.K. Sfantos and M.H. Aliabadi، نويسنده , , M.H. Aliabadi، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2007
Pages
12
From page
378
To page
389
Abstract
In this paper an application of the boundary element method for simulating wear in total hip prosthesis is presented. Several examples including different update periods of the worn acetabular cup, various femoral head sizes and various materials for both the femoral head and the acetabular cup are simulated under the same variable loading conditions for up to 20 years of service. Moreover, two different femoral models are considered in order to investigate the influence of the femoral modelling. The analysis demonstrates that due to the boundary only modelling requirement, the computational time and storage remains low, allowing large service periods to be simulated. Generally, the results obtained are in good agreement with other researchers findings. Moreover, ignoring the bending of the femoral neck in the model, results in a small overestimation of the maximum wear depth, while the volumetric wear is slightly underestimated. However, these differences are trivial considering the reduction of the computational effort.
Keywords
Total hip arthroplasty , Wear simulation , Boundary element method
Journal title
Journal of Biomechanics
Serial Year
2007
Journal title
Journal of Biomechanics
Record number
452432
Link To Document