Title of article :
Comparison of the hemodynamics in 6 mm and 4–7 mm hemodialysis grafts by means of CFD
Author/Authors :
Ilse Van Tricht، نويسنده , , Dirk De Wachter، نويسنده , , Jan Tordoir، نويسنده , , Pascal Verdonck، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Abstract :
The aim of our study is to investigate with computational fluid dynamics (CFD) whether different arterial anastomotic geometries result in a different hemodynamics at the arterial (AA) and venous anastomosis (VA) of hemodialysis vascular access grafts. We have studied a 6 mm graft (CD) and a 4–7 mm graft (TG).
A validated three-dimensional CFD model is developed to simulate flow in the two graft types. Only the arterial anastomosis (AA) geometry differs. The boundary conditions applied are a periodic velocity signal at the arterial inlet and a periodic pressure wave at the venous outlet. Flow rate is set to 1000 ml/min. The time dependent Navier–Stokes equations are solved. Wall shear stress (WSS), wall shear stress gradient (WSSG) and pressure gradient (PG) are calculated.
Anastomotic flow is asymmetric although the anastomosis geometry is symmetric. The hemodynamic parameters, WSS, WSSG and PG, values at the suture line of the arterial anastomosis of the TG are at least twice as much as in the CD. Comparing the parameters at the two AA indicate that little flow rate increase introduces the risk of hemolysis in the TG whereas the CD is completely free of hemolysis. The hemodynamic parameter values at the venous anastomosis of the CD are 24 till 35% higher compared to the values of the TG.
WSS values (>3 Pa) in the VA are in the critical range for stenosis development in both graft geometries. The zones where the parameters reach extreme values correspond to the locations where intimal hyperplasia formation is reported in literature. In all anastomoses, the hemodynamic parameter levels are in the range where leucocytes and platelets get activated.
Our simulations confirm clinical results where TG did not show a better outcome when compared to the CD.
Keywords :
Hemodialysis , Vascular access , CFD
Journal title :
Journal of Biomechanics
Journal title :
Journal of Biomechanics