• Title of article

    Use of the site percutaneous breast biopsy device

  • Author/Authors

    E. R. E. Denton، نويسنده , , M. J. Mitchell ، نويسنده , , R. R. Nash، نويسنده , , Stephanie M. Bingham، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    3
  • From page
    107
  • To page
    109
  • Abstract
    A new technique is described for image guided percutaneous diagnostic excision biopsy of non-palpable mammographically detected breast lesions. This method involves insertion of a localizing guide wire under stereotactic guidance followed by insertion of the 15 mm diameter biopsy device which is positioned within the breast immediately proximal to the target lesion. A cylindrical cannula is advanced over the lesion which is then separated from the surrounding breast tissue by deploying a garrote wire at the distal end of the cannula. The biopsy device is then removed with the target lesion contained within the cannula. The mammographic lesion is excised without removing any excess adjacent breast tissue, and in one piece, so allowing histological examination of the whole lesion. The experience of the first three cases is described with imaging and histological correlation.
  • Journal title
    The Breast
  • Serial Year
    2000
  • Journal title
    The Breast
  • Record number

    454260