    Intra-operative specimen analysis using faxitron microradiography for excision of mammographically suspicious, non-palpable breast lesions

  • Author/Authors

    M. Muttalib، نويسنده , , M. Tisdall، نويسنده , , R. Scawn، نويسنده , , S. Shousha، نويسنده , , R. S. Cummins، نويسنده , , H. D. Sinnett، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    9
  • From page
    307
  • To page
    315
  • Abstract
    Recent advances in digital imaging have made Faxitron microradiography an attractive alternative to intra-operative conventional specimen radiography (CSR) for the excision of wire-localized breast lesions. Faxitron specimen analysis time, usefulness of digital image manipulation and re-excision rates were evaluated in comparison to CSR in 299 consecutive wire-localized excisions for mammographically suspicious non-palpable breast lesions (172 procedures with Faxitron, 127 with CSR) in a non-randomized study. The corresponding mean operation times were 34.7 vs. 42.7 min and the respective re-excision rates were 19.8% vs. 31.5% (no significant difference on χ2 analysis P<0.1). Faxitron digital image manipulation led to cavity biopsies in 50% (60/121) of the cancer excisions. In 19 of these (16%), histological excision margins were converted from incomplete to complete. The shorter Faxitron mean operating time enables an additional wire-localized operation per theatre list. Digital imaging guides the surgeon for additional cavity biopsies, resulting in re-excision rates as good as CSR.