Title of article :
Endoscopic third ventriculostomy in the treatment of shunt-related over-drainage: Preliminary experience with a new approach how to render ventricles navigable
Author/Authors :
Jürgen M. Boschert، نويسنده , , Joachim K. Krauss، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
Objective
Endoscopic third ventriculostomy (ETV) is increasingly used for the treatment of shunt-related complications in hydrocephalic patients, particularly if the etiology of the underlying hydrocephalus is of obstructive nature.
Purpose
Due to the slit-like configuration of the ventricles, ventricular dilatation must be achieved prior to ETV in patients with problems due to over-drainage. This has been accomplished by ligating or explanting the shunt. Here we present an alternative procedure using a gravitational antisiphon device.
Materials and methods
In two patients with over-drainage due to shunted occlusive hydrocephalus ventricles were dilated by integrating an antisiphon device (Miethke ShuntAssistant, Aesculap AG, Tuttlingen, Germany) into their shunt systems. The resistance of the antisiphon device, which is available in configurations from 10 to 35 cm H2O was chosen 10 cm H2O higher than necessary to prevent siphoning in the individual patient.
Results
Both patients gradually recovered from their over-drainage symptoms and the ventricles enlarged enough to allow access with an endoscope. Using a standard procedure, ETV was performed 7 days and 1 month later, respectively. In the same operative session the shunts were occluded. Shunts were removed within 6 weeks after ETV. During follow-up of more than 3 years, both patients remained free of symptoms.
Conclusion
The incorporation of an antisiphon device with resistance level selected 10 cm H2O higher than needed to prevent anti-siphoning into a pre-existing shunt system in patients suffering from shunt-related over-drainage is a safe and effective technique to render ventricles large enough to allow endoscopic access for ETV.
Keywords :
hydrocephalus , Shunt , Endoscopic third ventriculostomy , Over-drainage , Slit-ventricle syndrome
Journal title :
Clinical Neurology and Neurosurgery
Journal title :
Clinical Neurology and Neurosurgery