Adrian S.H. Cheng، نويسنده , , Tammy J. Pegg، نويسنده , , Theodoros D. Karamitsos، نويسنده , , Nick Searle، نويسنده , , Michael Jerosch-Herold، نويسنده , , Robin P. Choudhury، نويسنده , , Adrian P. Banning، نويسنده , , Stefan Neubauer، نويسنده , , Matthew D. Robson، نويسنده , , Joseph B. Selvanayagam، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
Objectives
This study was designed to establish the diagnostic accuracy of cardiovascular magnetic resonance (CMR) perfusion imaging at 3-Tesla (T) in suspected coronary artery disease (CAD).
Background
Myocardial perfusion imaging is considered one of the most compelling applications for CMR at 3-T. The 3-T systems provide increased signal-to-noise ratio and contrast enhancement (compared with 1.5-T), which can potentially improve spatial resolution and image quality.
Methods
Sixty-one patients (age 64 ± 8 years) referred for elective diagnostic coronary angiography (CA) for investigation of exertional chest pain were studied (before angiogram) with first-pass perfusion CMR at both 1.5- and 3-T and at stress (140 μg/kg/min intravenous adenosine, Adenoscan, Sanofi-Synthelabo, Guildford, United Kingdom) and rest. Four short-axis images were acquired during every heartbeat using a saturation recovery fast-gradient echo sequence and 0.04 mmol/kg Gd-DTPA bolus injection. Quantitative CA served as the reference standard. Perfusion deficits were interpreted visually by 2 blinded observers. We defined CAD angiographically as the presence of ≥1 stenosis of ≥50% diameter in any of the main epicardial coronary arteries or their branches with a diameter of ≥2 mm.
Results
The prevalence of CAD was 66%. All perfusion images were found to be visually interpretable for diagnosis. We found that 3-T CMR perfusion imaging provided a higher diagnostic accuracy (90% vs. 82%), sensitivity (98% vs. 90%), specificity (76% vs. 67%), positive predictive value (89% vs. 84%), and negative predictive value (94% vs. 78%) for detection of significant coronary stenoses compared with 1.5-T. The diagnostic performance of 3-T perfusion imaging was significantly greater than that of 1.5-T in identifying both single-vessel disease (area under receiver-operator characteristic [ROC] curve: 0.89 ± 0.05 vs. 0.70 ± 0.08; p < 0.05) and multivessel disease (area under ROC curve: 0.95 ± 0.03 vs. 0.82 ± 0.06; p < 0.05). There was no difference between field strengths for the overall detection of coronary disease (area under ROC curve: 0.87 ± 0.05 vs. 0.78 ± 0.06; p = 0.23).
Conclusions
Our study showed that 3-T CMR perfusion imaging is superior to 1.5-T for prediction of significant single- and multi-vessel coronary disease, and 3-T may become the preferred CMR field strength for myocardial perfusion assessment in clinical practice.
Keywords :
myocardial infarction , PET , RCA , SPECT , SNR , positron emission tomography , Receiver operating characteristic , Left ventricle , intravenous , coronary artery disease , ROC , Signal-to-noise ratio , LAD , Single-photon emission computed tomography , MI , CAD , ROI , T , Region of interest , IV , Signal intensity , Si , CMR , Right coronary artery , LV , left anterior descending , cardiovascular magnetic resonance , CNR , contrast-to-noise ratio , Tesla