Transthoracic defibrillation: Importance of avoiding electrode placement directly on the female breast
Author/Authors :
Luis A. Pagan-Carlo، نويسنده , , Kirk T. Spencer، نويسنده , , Colin E. Robertson، نويسنده , , Anne Dengler، نويسنده , , Clay Birkett، نويسنده , , Richard E. Kerber، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1996
Pages :
4
From page :
449
To page :
452
Abstract :
Objectives. This study sought to determine the effect on transthoracic Impedance of placement of defibrillation electrodes on the female breast versus adjacent to or under the breast. Background. Transthoracic impedance is major determinant of transthoracic current flow in defibrillation. For given energy setting, high transthoracic impedance reduces current flow and may adversely affect the ability of electric shocks to accomplish defibrillation. We hypothesized that the increased interelectrode tissue associated with placement of the apex defibrillation electrode on the female breast would result in increased transthoracic impedance compared with electrode placement lateral to or under the breast. Method. Transthoracic impedance was measured noninvasively by passing 5-V, 31.25 kHz square wave current through the chest and comparing the low level current flow to known references. We measured transthoracic impedance associated with three different apex defibrillation electrode positions—on the breast, under the breast and lateral to the breast—in 25 women (brassiere size 34 to 48C 65 to 75 years old, body weight 128 to 328 lb [58 to 148 kg] and 2 mm. The measurements were taken with modified defibrillator that accurately predicts transthoracic impedance without delivering an actual shock. The measurement sequence was random. Results. The average measured transthoracic impedance with placement of the apex defibrillation electrode on the breast was 95 ± 25 ohms (mean ± SD), under the breast 84 ± 17* ohms and lateral to the breast 83 ± 20* ohms (asterisk indicates p < 0.01 vs. on the breast by analysis of variance). The study cohort was also classified into two groups: large breasted (brassiere size ≥ 40) and small breasted (brassiere size ≤39). The measured transthoracic impedances for the large-breasted group were 112 ± 20 ohms for on the breast, 94 ± 13* ohms for under the breast and 98 ± 19* ohms for lateral to the breast. For the small breasted group, the similar transthoracic impedance measurements were 81 ± 21, 77 ± 16 and 71 ± 13* ohms, respectively. Conclusions. In women, placement of the apex defibrillation electrode on the breasts results in higher transthoracic impedance, which will reduce current flow. We recommend placing the apex electrode lateral to or underneath the breast.
Journal title :
JACC (Journal of the American College of Cardiology)
Serial Year :
1996
Journal title :
JACC (Journal of the American College of Cardiology)
Record number :
479415
Link To Document :
بازگشت