Balloon angioplasty for the treatment of coronary in-stent restenosis: immediate results and 6-month angiographic recurrent restenosis rate
Author/Authors
Hélène Eltchaninoff، نويسنده , , René Koning MD، نويسنده , , Christophe Tron، نويسنده , , Vivek Gupta، نويسنده , , Alain Cribier، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 1998
Pages
5
From page
980
To page
984
Abstract
Objectives. The purpose of this prospective study was to evaluate the immediate results and the 6-month angiographic recurrent restenosis rate after balloon angioplasty for in-stent restenosis.
Background. Despite excellent immediate and mid-term results, 20% to 30% of patients with coronary stent implantation will present an angiographic restenosis and may require additional treatment. The optimal treatment for in-stent restenosis is still unclear.
Methods. Quantitative coronary angiography (QCA) analyses were performed before and after stent implantation, before and after balloon angioplasty for in-stent restenosis and on 6-month systematic coronary angiogram to assess the recurrent angiographic restenosis rate.
Results. Balloon angioplasty was performed in 52 patients presenting in-stent restenosis. In-stent restenosis was either diffuse (≥ 10 mm) inside the stent (71%) or focal (29%). Mean stent length was 16 ± 7 mm. Balloon diameter of 2.98 ± 0.37 mm and maximal inflation pressure of 10 ± 3 atm were used for balloon angioplasty. Angiographic success rate was 100% without any complication. Acute gain was lower after balloon angioplasty for in-stent restenosis than after stent implantation: 1.19 ± 0.60 mm vs. 1.75 ± 0.68 mm (p = 0.0002). At 6-month follow-up, 60% of patients were asymptomatic and no patient died. Eighteen patients (35%) had repeat target vessel revascularization. Angiographic restenosis rate was 54%. Recurrent restenosis rate was higher when in-stent restenosis was diffuse: 63% vs. 31% when focal, p = 0.046.
Conclusions. Although balloon angioplasty for in-stent restenosis can be safely and successfully performed, it leads to less immediate stenosis improvement than at time of stent implantation and carries high recurrent angiographic restenosis rate at 6 months, in particular in diffuse in-stent restenosis lesions.