Title of article :
Ventricular tachycardi during follow-up in patients resuscitated from ventricular fibrillation: experience from stored electrograms of implantable cardioverter-defibrillators
Author/Authors :
Rudolf Rüppel، نويسنده , , Catarin A. Schlüter، نويسنده , , Sigrid Boczor، نويسنده , , Thomas Meinertz، نويسنده , , Michael Schlüter، نويسنده , , Karl-Heinz Kuck، نويسنده , , Riccardo Cappato، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1998
Pages :
7
From page :
1724
To page :
1730
Abstract :
Objective. The purpose of this study was to use the electrogram storage capabilities of the implantable cardioverter-defibrillator (ICD) to categorize any arrhythmic event during follow-up in group of patients who had survived an episode of ventricular fibrillation (VF) and to possibly identify clinical predictors of future arrhythmic events. Background. Little is known about the electrophysiologic characteristics of ventricular arrhythmias recurring during follow-up in survivors of VF as the sole documented arrhythmi at the time of resuscitation. Methods. Forty patients (58 ± 10 years; 73% men; left ventricular ejection fraction 42 ± 18%; 70% with coronary artery disease) who had survived an episode of VF and subsequently received an ICD capable of intracardiac electrogram recording and storage were followed for 23 ± 11 months. In all patients, the arrhythmogenic substrate was investigated by means of programmed electrical stimulation (PES). Results. Among the 40 patients, 41 episodes of ventricular arrhythmias were documented in 13 patients (33%): 36 episodes of ventricular tachycardias (VT) were recorded in 11 patients (28%) and 5 episodes of VF were recorded in the remaining 2 patients (5%). Age, gender, cardiac disease and left ventricular ejection fraction failed to distinguish between patients with clinical recurrences and patients without. The sensitivity, specificity and positive accuracy of PES were 29%, 63% and 46%, respectively, for prediction of ventricular arrhythmi recurrence; 45%, 70% and 36%, respectively, for prediction of VT; and 50%, 98% and 50%, respectively, for prediction of VF during follow-up. Conclusions. In survivors of VF receiving ICD therapy, VT is the most common ventricular arrhythmi recorded on device-incorporated electrograms during follow-up. This finding, associated with the relatively well-preserved ventricular function, may account for the ability of these patients to survive at time of the index arrhythmia; the use of antitachycardi pacing as modality to treat arrhythmi recurrences may contribute to reduce the incidence of shock during follow-up in these patients.
Keywords :
CAD , coronary artery disease , Ventricular tachycardia , ICD , PES , Vf , ventricular fibrillation , VT , atrioventricular , AV , implantable cardioverter-defibrillator , programmed electrical stimulation
Journal title :
JACC (Journal of the American College of Cardiology)
Serial Year :
1998
Journal title :
JACC (Journal of the American College of Cardiology)
Record number :
480933
Link To Document :
بازگشت