Title of article :
Specific Heparin Preparations Interfere With the Simultaneous Measurement of Ionized Magnesium and Ionized Calcium
Author/Authors :
Martha E. Lyon، نويسنده , , Debbie Bremner، نويسنده , , Thomas Laha، نويسنده , , Sohail Malik، نويسنده , , Paul J. Henderson، نويسنده , , Margaret A. Kenny، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 1995
Pages :
6
From page :
79
To page :
84
Abstract :
Objective: To determine whether heparin anticoagulants used for analysis of whole blood ionized calcium would influence the measurement of ionized magnesium. Methods: The effects of zinc heparin, lithium heparin, and electrolyte-balanced heparin on the simultaneous measurement of ionized magnesium and ionized calcium in serum were determined using ion selective electrodes. Results: Time-dependent biases in ionized magnesium and calcium concentrations were apparent with zinc heparin but not with lithium or electrolyte-balanced heparin. Ionized magnesium and calcium concentrations were more significantly influenced by volume-dependent changes in zinc heparin potency than with lithium or electrolyte-balanced heparin. Conclusion: Zinc heparin produces a significant positive bias in the simultaneous determination of ionized magnesium and ionized calcium concentrations
Keywords :
analysis , ionized calcium , Serum. , ionized magnesium , zinc heparin , lithium heparin , electrolyte-balanced heparin , preanalytical error
Journal title :
Clinical Biochemistry
Serial Year :
1995
Journal title :
Clinical Biochemistry
Record number :
481396
Link To Document :
بازگشت