• Title of article

    Evaluation of the Technicon Immuno I® Random Access Immunoassay Analyzer and Calculation of Pediatric Reference Ranges for Endocrine Tests, T-Uptake, and Ferritin

  • Author/Authors

    Jayasimha N. Murthy، نويسنده , , Jocelyn M. Hicks، نويسنده , , Steven J. Soldin، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 1995
  • Pages
    5
  • From page
    181
  • To page
    185
  • Abstract
    Objectives: Evaluation of the precision, accuracy, and user-friendliness of the Technician Immuno I. Calculation of pediatric reference ranges for ferritin and endocrine tests run on Immuno I. Methods: Precision and accuracy were measured using controls and method comparison studies. Pediatric reference ranges were calculated by comparing the Immuno I results for 100 patients with those of the Abbott IMx and TDx. The regression equation obtained was then used to convert the IMx and TDx reference ranges to reference ranges for the Immuno I. Results: The Immuno I provided both accurate and precise measurement of drugs and endocrine hormones. Pediatric reference ranges were obtained for ferritin and all endocrine tests. Conclusion: The Immuno I is user-friendly and provides reliable measurement of both the drugs tested and endocrine tests on a micro-sample. Reagent and curve stability are excellent.
  • Keywords
    Correlation , Precision , Immunoassay , Technicon Immuno I® analyzer , Abbott TDx and IMx , pediatricreference ranges for hormones.
  • Journal title
    Clinical Biochemistry
  • Serial Year
    1995