• Title of article

    Analytical evaluation of the Bio-Rad Variant II automated HbA1C analyzer

  • Author/Authors

    Trefor N. Higgins، نويسنده , , George B. Blakney، نويسنده , , Jan Dayton، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    5
  • From page
    361
  • To page
    365
  • Abstract
    Objectives: To evaluate the analytical performance of the Bio-Rad Variant II HbA1C analyzer (a completely automated system for the quantification of glycohemoglobin [HbA1C] in blood). Design and Methods: The analytical parameters of precision, linearity and analytical range were assessed and HbA1C results from the Variant II were compared to HbA1C results from the Bio-Rad Variant (a method certified by the National Glycohemoglobin Standardization Program). The effect of a variety of hemoglobin variants on HbA1C obtained on the system was investigated. Results: Total imprecision was less than 5% and the results compared well with those from an established method. The method has a wide analytical range with no carryover between specimens. Conclusion: The HbA1C method on the Variant II gives acceptable analytical performance.
  • Journal title
    Clinical Biochemistry
  • Serial Year
    2001
  • Journal title
    Clinical Biochemistry
  • Record number

    482216