• Title of article

    Determination of myoglobin: comparative evaluation of the new automated VIDASR assay with two other immunoassays

  • Author/Authors

    Françoise Le Moigne، نويسنده , , Marie-Christine Beauvieux، نويسنده , , Philippe Derache، نويسنده , , Yves-Michel Darmon، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    8
  • From page
    255
  • To page
    262
  • Abstract
    Objectives: Myoglobin provides the earliest indication of acute myocardial infarction. In this study, the new myoglobin assay for the VIDAS system (bioMérieux) was evaluated. Design and methods: This assay, using an enzyme-linked fluorescent immunoassay (ELFIA) method, was compared with the OlympusR immunoturbidimetric method and with another immunometric method (Immulite turboR) using an enzyme-linked chemiluminescent immunoassay (CLIA). Results: The CVs for within-and between-run reproducibility are very similar for the tested methods and acceptable linearity ranges were obtained. No significant interference of hemolysis, turbidity and icteria was observed. In the whole cohort, we obtained decreased values over the entire range of the assay with VIDAS and Immulite turbo methods compared to the Olympus assay; this is probably mainly linked to differences in standards used due to the absence of international standardization of the myoglobin determination. Conclusions: The new VIDAS myoglobin automated assay provides biologists with a rapid, accurate and reliable determination of myoglobin in plasma samples collected during cardiac workup.
  • Keywords
    chemiluminescence , Immunoassay , Turbidimetry , Fluorimetry , myoglobin
  • Journal title
    Clinical Biochemistry
  • Serial Year
    2002
  • Journal title
    Clinical Biochemistry
  • Record number

    482300