• Title of article

    Analytical evaluation of hemoglobin A1c dual kit assay on Bio-Rad Variant II: an automated HPLC hemoglobin analyzer for the management of diabetic patients

  • Author/Authors

    Majid Y. Moridani، نويسنده , , Zul Verjee، نويسنده , , Lynn C. Allen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    4
  • From page
    317
  • To page
    320
  • Abstract
    Objectives To evaluate the analytical performance of the Bio-Rad Variant II HbA1c dual kit assay. Design and methods Precision, carryover, linearity and analytical range were investigated. 139 patients’ HbA1c results analyzed by the Variant II were compared to the Variant I method. 49 blood samples analyzed by the Variant II at Toronto Medical Laboratories (TML) were compared to the Variant II at Hospital for Sick Children (HSC). Results Total imprecision was less than 2% for the Variant II assay. The method had a wide analytical range with no carryover. HbA1c results were not changed after switching back and forth from the beta thalassemia to HbA1c assay. The Variant II showed an average of 0.0027 negative bias compared to the Variant I method. There was an average of 0.0020 negative bias for HbA1c results on the Variant II at TML compared to the Variant II at HSC. Conclusion HbA1c analysis on the Variant II HbA1c dual kit is a relatively fast and reproducible method.
  • Keywords
    Diabetes , Hemoglobin A1c , HbA1C , Glycated Hb , Variant II , Bio-Rad Variant II Dual Kit
  • Journal title
    Clinical Biochemistry
  • Serial Year
    2003
  • Journal title
    Clinical Biochemistry
  • Record number

    482414