    Evaluation of the analytic performances of the new HPLC system HLC-723 G7 for the measurement of hemoglobin A1C

  • Author/Authors

    Alessandro Terreni، نويسنده , , Renata Paleari، نويسنده , , Anna Caldini، نويسنده , , Agostino Ognibene، نويسنده , , Andrea Mosca، نويسنده , , Gianni Messeri، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    4
  • From page
    607
  • To page
    610
  • Abstract
    Objectives The analytical performance of a new automated HPLC system, for the determination of HbA1C in blood (Tosoh HLC-723 G7), was studied. Design and methods The study design included the evaluation of imprecision, linearity, interference and carryover. Comparison study was performed by comparing HbA1C results with those obtained from an established method (Bio-Rad Variant II). Results Total imprecision was less than 1.34% and the results were linear up to 17.2% HbA1C. The method showed a wide analytical range, and no carryover between specimens. Comparison study yielded, R = 0.989, Sy.x = 0.255, regression equation (y = 0.9895x−0.35); Bland–Altman plot showed a mean BIAS = – 0.43% HbA1C with confidence limits ranging from −0.48% to −0.38% HbA1C. The presence of abnormal hemoglobin was clearly revealed, and no interference from labile HbA1C was apparent. Conclusion The HLC-723 G7 instrument seems to be a reliable system for routine assay of HbA1C.
  • Keywords
    evaluation , HPLC , HbA1C , glycohemoglobin