Title of article :
High-performance liquid chromatographic method with column-switching and post-column reaction for determination of serotonin levels in platelet-poor plasma
Author/Authors :
Yuji Hirowatari، نويسنده , , Katsuko Hara، نويسنده , , Hiroshi Kamihata، نويسنده , , Toshiji Iwasaka، نويسنده , , Hakuo Takahashi، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2004
Abstract :
Objectives: Measurement of serotonin concentrations in blood is considered to be important for the clinical studies of diseases involving vascular endothelial injury. We have developed a high-performance liquid chromatographic (HPLC) method for measurement of serotonin level in platelet-poor plasma (PPP).
Methods: Venous blood samples were collected using vacuum tubes containing ethylenediaminetetraacetic acid (EDTA) dipotassium salt; the optimum concentration in blood was 7.4 mmol/l. Serotonin in samples was separated by HPLC with a column-switching system, and was specifically converted into a fluorescent derivative with benzylamine for convenient detection.
Results: The between-day assay coefficient of variation for a serotonin level (5.1 nmol/l) in PPP was 6.2%. The mean concentration of serotonin in PPP in healthy subjects was 5.7 ± 3.0 nmol/l. The serotonin levels in PPP in patients with ischemic heart disease were significantly higher than those in healthy subjects (P = 0.001).
Conclusions: The developed method had sufficient performance to determine serotonin levels in PPP from patients with ischemic heart disease.
Keywords :
chromatography , Serotonin , platelet-poor plasma
Journal title :
Clinical Biochemistry
Journal title :
Clinical Biochemistry