Title of article :
Validation of 5 routine assays for serum free testosterone with a candidate reference measurement procedure based on ultrafiltration and isotope dilution–gas chromatography–mass spectrometry
Author/Authors :
K. Van Uytfanghe، نويسنده , , D. St?ckl، نويسنده , , J.M. Kaufman، نويسنده , , T. Fiers، نويسنده , , A. De Leenheer، نويسنده , , L.M. Thienpont، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
Background/Method:
The analytical validity of free testosterone (FTe) analog immunoassays is subject to much controversy. We revisited the validation of 4 analog assays and 1 FTe calculation procedure with a metrologically traceable reference measurement procedure (RMP) based on ultrafiltration and isotope dilution–mass spectrometry for direct measurement of Te in the ultrafiltrate. To this end, we performed split-sample measurements of 40 male sera.
Results:
Deming regression showed that 3 of the immunoassays had moderate to good correlation (0.8474 ≤ r ≤ 0.9241) with the RMP; however, the slope was markedly below 1. The FTe calculation procedure was in good agreement with this result. The Sy/x values for all assays were higher than the combined imprecision values, which indicate their susceptibility to matrix-related effects.
Conclusions:
The study demonstrated substantial differences in analytical quality of FTe assays; however, the results suggested that after extending the validation with a larger variety of samples, recalibration of some analog assays might be worth further investigation.
Keywords :
ultrafiltration , Serum free testosterone , Reference measurement procedure , Isotope dilution–gas chromatography–mass spectrometry , Methodcomparison