    Improved infrared spectrophotometer for point-of-care patient 13C-urea breath testing in the primary care setting

  • Author/Authors

    Antone R. Opekun، نويسنده , , Ann B. Gotschall، نويسنده , , Nageeb Abdalla، نويسنده , , Cornelia Agent، نويسنده , , Elizabeth Torres، نويسنده , , Fred M. Sutton، نويسنده , , David Y. Graham، نويسنده , , Kyoko Tsuchiya، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2005
  • Pages
    4
  • From page
    731
  • To page
    734
  • Abstract
    The 13C-urea breath test provides non-invasive testing for Helicobacter pylori infection with the possibility of analysis at the point of care. Point of care tests require accurate and efficient desktop instrumentation. Objective: To compare results obtained from a new infrared spectrophotometer (10 kg, 2 min sample measurement) to the previously approved UBiT®-IR300 (22.5 kg, 5–6 min sample measurement). Design and methods: Adults underwent urea breath testing; breath samples were analyzed using the new and the predicate instruments. Results: There were 220 adults enrolled (age range 18–74 years; M/F = 35:65); with 86 UBT positive and 134 negative cases. The overall agreement between instruments was 99.6% (95% C.I. = 97.6 to 99.9); the positive agreement was 100%. Conclusion: Technical performance of the new instrument was excellent.
  • Keywords
    Helicobacter pylori , urea , clinical trial , Breath testing , Infrared spectrophotometer