Title of article :
Evaluation of HbA1c on the Roche COBAS Integra 800 closed tube system
Author/Authors :
James K. Fleming، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Pages :
6
From page :
822
To page :
827
Abstract :
Objectives: Evaluate a new whole blood (WB) HbA1c immunoassay and system with closed tube sampling (CTS) capability. Design and methods: Compare the Tina-quant Haemoglobin A1c Gen.2 (A1C-2) application on the COBAS INTEGRA 800 (I800) and new I800 dedicated system with CTS capability to current Integra applications and a HbA1c method accurate with common haemoglobin (Hb) variants. Results: CVs were ≤ 1.7%. Mean bias against National Glycohaemoglobin Standardization Program (NGSP) samples was 0.3 HbA1c %. Compared to the Hitachi Tina-quant® [a] HbA1c II (HbA1c II) assay (accurate with common Hb variants), mean bias was 0.04% and 0.21% HbA1c at 6% and 9%, respectively, with Hb AS variants; and − 0.01% and 0.26% HbA1c at 6% and 9%, respectively, with Hb AC variants. Conclusions: The Integra A1C-2 application is precise, accurate against NGSP-assigned samples and the Hb variants tested; and, the I800 dedicated system with CTS capability offers increased throughput and reduced sample handling.
Keywords :
HbA1C , Closed tube sampling (CTS) , COBAS Integra® 800 (I800) , COBAS Integra Haemoglobin A1C Gen.2 (A1C-2) , Haemoglobin (Hb) , National Glycohaemoglobin Standardization Program (NGSP) , Secondary Reference Laboratory (SRL) , Haemoglobin variants , Glycatedhaemoglobin
Journal title :
Clinical Biochemistry
Serial Year :
2007
Journal title :
Clinical Biochemistry
Record number :
484994
Link To Document :
بازگشت