Title of article :
QA aspects for HbA1c measurements
Author/Authors :
Trefor N. Higgins، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
3
From page :
88
To page :
90
Abstract :
Objectives: To provide mechanisms for evaluating HbA1c results that meet the criteria for review by the 2002 NACB guidelines for reporting HbA1c values. Design and methods: Complete blood count (CBC) data and comparison of obtained HbA1c with a calculated HbA1c were used to assess the validity of HbA1c results meeting the NACB review criteria. Results: The use of CBC data and a calculated HbA1c were found to be useful in evaluating the validity of unusual HbA1c results. Conclusions: The validity of high and low HbA1c results can be checked by the review of CBC data and comparing a calculated HbA1c against the measured value.
Keywords :
HBA1C , quality assurance , Glycohemoglobin
Journal title :
Clinical Biochemistry
Serial Year :
2008
Journal title :
Clinical Biochemistry
Record number :
485112
Link To Document :
بازگشت