Title of article :
Hitachi Hemolytic Index correlates with HBOC-201 concentrations: Impact on suppression of analyte results
Author/Authors :
Paula F. Moon-Massat، نويسنده , , James P. Tierney، نويسنده , , Karl G. Hock، نويسنده , , Mitchell G. Scott، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
4
From page :
432
To page :
435
Abstract :
Objectives To determine the feasibility of laboratories to use an instrumentʹs Hemolytic Index (HI) to determine if a test result would be accurate in the presence of hemoglobin-based oxygen carrier HBOC-201. Design and methods HI values from the Roche Hitachi Modular P800 for samples containing HBOC-201 were determined. HI limits for 24 tests determined by addition of RBC lysate hemoglobin to serum and the instrument manufacturerʹs HI limits were compared to HI limits for reporting the same tests determined by adding HBOC-201 to plasma. Results There is a linear relationship (R2 = 0.99) between the HI value on the Modular P800 and HBOC-201 concentrations. Twenty-one of 24 HI limits for reporting results as determined by the manufacturer or adding RBC lysate to serum were more conservative or equal to the limit determined by HBOC-201 interference testing. HBOC-201 interference testing was more conservative for albumin, creatinine, and uric acid. Conclusion For most analytes, the Modular P800 HI provides an accurate, conservative and automated method to determine whether laboratory results should be reported or suppressed when samples contain HBOC-201.
Keywords :
interference , Hemolysis , Hemoglobin based oxygen carrier , Blood substitute , Oxygen therapeutic
Journal title :
Clinical Biochemistry
Serial Year :
2008
Journal title :
Clinical Biochemistry
Record number :
485164
Link To Document :
بازگشت