Title of article :
Performance evaluation of ion exchange and affinity chromatography for HbA1c estimation in diabetic patients with HbD: A study of 129 samples
Author/Authors :
Bushra Moiz، نويسنده , , Mashhooda Rasool Hashmi، نويسنده , , Shahmina Sadaf، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Abstract :
Objectives
To compare ion exchange and boronate affinity chromatography for HbA1c estimation in patients with type I and II diabetes having hemoglobin D.
Design and methods
Systems based on ion exchange and boronate affinity chromatography were evaluated and compared for their performance for HbA1c estimation in patients with homozygous and heterozygous D disease.
Results
Boronate affinity chromatography shows least interference by HbD in heterozygous as well as homozygous diabetic patients for HbA1c estimation.
Conclusions
The use of boronate affinity chromatography was found to be helpful in evaluating glycemic control in diabetic subjects with HbD.
Keywords :
HBA1C , Diabetes , chromatography , HbD , Hemoglobin variant
Journal title :
Clinical Biochemistry
Journal title :
Clinical Biochemistry