Development and validation of a fast quantitative method for plasma dimethylarginines analysis using liquid chromatography–tandem mass spectrometry
Author/Authors
Oceania DʹApolito، نويسنده , , Giuseppe Paglia، نويسنده , , Filomena Tricarico، نويسنده , , Daniela Garofalo، نويسنده , , Alessandra Pilotti، نويسنده , , Olga Lamacchia، نويسنده , , Mauro Cignarelli، نويسنده , , Gaetano Corso، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
5
From page
1391
To page
1395
Abstract
Objectives
The aim of this work was to implement a fast, accurate and simple method to quantify plasma ADMA and SDMA, in a run time suitable for routine analysis.
Design and methods
We developed and validated a hydrophilic interaction chromatographic method coupled to tandem mass spectrometry (HILIC–MS/MS) for separation and simultaneous quantification of Arginine (Arg) and its dimethylarginines, ADMA and SDMA, with a short run time (less than 5 min) using a small volume of human plasma (0.02 mL).
Results
Correlation coefficients (r) of the calibration curves ranged from 0.9926 to 0.9984. Within-day and between-day imprecision (CV%) and inaccuracy (%), carry-over and recovery were also evaluated for validation. Preliminary data of Arg, ADMA and SDMA from 30 apparently healthy subjects and type 2 diabetic patients (n = 33) with and without kidney dysfunction were calculated and some statistical differences occurred among them (p < 0.05).
Conclusions
Data from calibration curves and quality controls reveal that the method is accurate and precise. Healthy subjects and diabetic patientsʹ values are in agreement with those reported in other studies.
Keywords
type 2 diabetes , L-Arginine , HILIC–MS/MS , Asymmetric dimethylarginine (ADMA) , Symmetric dimethylarginine (SDMA)